Description of the MCA527 Firmware Commands
Command name
CMD_QUERY_POWER (Continuation)
Result data array
Byte offset 72 +12V primary current [mA] at stop unsigned long
Byte offset 76 -12V primary current [mA] at stop unsigned long
Byte offset 80 +24V primary current [mA] at stop unsigned long
Byte offset 84 -24V primary current [mA] at stop unsigned long
Byte offset 88 Battery voltage [mV] at stop unsigned long At the MCA527Micro this data
corresponds to the USB input
voltage.
Byte offset 92 HV [ * 1.2 V] at stop unsigned long
Byte offset 96 ADC correction offset on SUB-D9 pin3 [LSB] char
Byte offset 97 Gain correction factor on SUB-D9 pin3 char factor = 0.001 * value + 1
Byte offset 98 unused 2 bytes
Byte offset 100 +12V actual value [ * 0.0625 V] at stop unsigned char
Byte offset 101 -12V actual value [ * 0.0625 V] at stop unsigned char
Byte offset 102 +24V actual value [ * 0.125 V] at stop unsigned char
Byte offset 103 -24V actual value [ * 0.125 V] at stop unsigned char
Byte offset 104 Voltage on SUB-D9 pin3 [ * 0.3125 mV] at stop unsigned short
Byte offset 106 Command flag and parameters 8 bytes
Byte offset 114 Voltage on SUB-D9 pin5 [ * 0.3125 mV] at stop unsigned short
Byte offset 116 Charger current [mA] at stop unsigned long
Byte offset 120 unused 6 bytes
Byte offset 126 Checksum unsigned short
Byte offset 128 MCA state integer value
STATE_READY = 1
STATE_RUN = 2
STATE_SUSPEND = 3
STATE_FINISH = 4
STATE_STOP = 5
STATE_FAIL = 6
STATE_WAIT_FOR_TRIGGER = 7
Byte offset 130 unused 2 bytes